Able are AMETEK Brookfield’s Exclusive Partner in the UK and Ireland for the AMETEK Brookfield Jerome J605 Fixed Monitoring Hydrogen Sulphide Analyser and can assist you and your customers in the following areas:
- Training in good practice in use and maintenance
- Offer Hire and Loan instrument service
- Assistance with Surveys including Offshore
- Portable or Fixed Monitoring variants available
Further to the above, Able can offer custom solutions based on your individual requirements. Email us now at info@ableinstruments.com
AMETEK Brookfield Jerome J605 Fixed Monitoring Hydrogen Sulphide Analyser
AMETEK Brookfield Gold Film Sensor Technology in a Weatherproof Enclosure for Continuous Hydrogen Sulphide Monitoring
Able Instruments are AMETEK Brookfield’s Exclusive Partner in the UK and Ireland for the AMETEK Brookfield Jerome J605 Fixed Monitoring Hydrogen Sulphide Analyser. The Jerome J605 Fixed Monitoring Hydrogen Sulphide Analyser is designed for continuous, long-term monitoring of hydrogen sulphide in demanding environmental conditions. The system utilises AMETEK Brookfield’s highly specific, patented gold film sensor technology to provide reliable trace level H₂S detection for safety, environmental and occupational exposure monitoring.
Engineered for fenceline and perimeter monitoring applications, the analyser is supplied within a robust GRP environmental enclosure rated to IP66/7. This allows permanent outdoor installation where adverse weather, temperature extremes and site conditions would prevent the use of standard portable instruments. Integrated temperature control supports stable internal operating conditions for reliable long-term performance.
Key Features:
- Gold film sensor technology: Patented sensor design for highly specific, trace level hydrogen sulphide detection.
- Fixed perimeter monitoring: Designed for continuous, long-term H₂S monitoring in fenceline and perimeter applications.
- Environmental enclosure: GRP polyester glass-reinforced enclosure rated to IP66/7 for outdoor installation.
- Integrated temperature control: Thermostatically controlled heating maintains stable internal operating conditions.
- Detachable handheld analyser: Jerome J605 analyser can be removed for survey work and source identification.
- Scalable site deployment: Multiple units can be installed to provide comprehensive perimeter coverage.
- Site system integration: Supports passive 4–20 mA output and digital connectivity for connection to monitoring or control systems.
- Project-ready options: Integrated data logging, remote viewing capability, mounting frames, commissioning and maintenance services available.
Have a particular project in mind? We’d welcome a call on +44 (0) 1189 311 188
The system can be mounted on a wall, pole, fence or custom-built frame, and multiple units may be deployed to create comprehensive perimeter monitoring coverage. An illuminated local display provides clear on-site visibility of operating status and measured values.
A key operational advantage is the detachable, handheld Jerome J605 analyser housed within the enclosure. The analyser can be removed for survey work, enabling operators to investigate elevated readings, identify H₂S sources and take corrective action without additional equipment.
The system supports digital and analogue outputs, including a passive 4–20 mA signal for integration with site monitoring or control systems. Optional data logging and remote viewing capability further support fixed monitoring applications where continuous hydrogen sulphide measurement is required.
Specifications:
| Detection Range | 3 ppb (0.003 ppm) to 10 ppm |
|---|---|
| Resolution | 20 ppt (0.02 ppb), range dependent |
| Accuracy / Precision | Range 0: ±1 ppb at 5 ppb / 10% RSD; Range 0: ±3 ppb at 50 ppb / 5% RSD; Range 1: ±0.03 ppm at 0.5 ppm / 5% RSD; Range 2: ±0.3 ppm at 5.0 ppm / 5% RSD |
| Response Time | Survey mode: 12 to 27 seconds; Manual mode: 12 to 27 seconds; Auto mode: 12 to 53 seconds |
| Flow Rate | 150 mL/min |
| Result Units | ppm, ppb |
| Autosample Intervals | 1, 2, 5, 10, 15, 30, 45, 60, 90 or 120 minutes |
| Integrated Data Storage | 20,000 test results; retains date, time and up to 80 locations |
| Display | 8-digit alphanumeric LCD |
| Battery | Rechargeable NiMH, 18-hour life, charges in 3 hours |
| Fuse | Auto-resetting fuse |
| Outputs | Digital USB serial data to PC, printer or USB flash drive; analogue 4–20 mA passive current loop, accurate to 0.3% of output |
| Operating Environment | -40 to +55 °C (temperature controlled enclosure) |
| Enclosure | Polyester glass-reinforced enclosure, IP66 rated |
| Dimensions | 600 (H) × 400 (W) × 260 mm (D) |
| Weight (loaded) | Approximately 12 kg |
| Mains Power Line Filter | 250 VAC, 50/60 Hz at 6 A |
| Switch Mode Power Supply | 24 VDC, 5 A, 120 W; 12 VDC, 1.5 A, 15 W |
| Internal Adjustable Thermostat | +5 to +60 °C, 10 A contact rating |