Roll over image to zoom in

Able are AMETEK Brookfield’s Exclusive Partner in the UK and Ireland and can assist you and your customers in the following areas:

  • Training in good practice in use and maintenance
  • Offer Hire and Loan instrument service
  • Assistance with Surveys including Offshore
  • Portable or Fixed Monitoring variants available

Further to the above, Able can offer custom solutions based on your individual requirements. Email us now at info@ableinstruments.com

AMETEK Brookfield Jerome J405 Fixed Monitoring Mercury Vapour Analyser

AMETEK Brookfield Gold Film Sensor Technology in a Weatherproof Enclosure for Continuous, Fixed Mercury Vapour Monitoring

Able Instruments are AMETEK Brookfield’s Exclusive Partner in the UK and Ireland for the AMETEK Brookfield Jerome J405 Fixed Monitoring Mercury Vapour Analyser. The Jerome J405 Fixed Monitoring Mercury Vapour Analyser is a heavy-duty system designed for continuous, long-term monitoring of airborne mercury vapour at trace concentration levels. The system utilises AMETEK Brookfield’s highly specific, patented gold film sensor for reliable detection in safety, environmental and occupational exposure monitoring applications.

Housed within a robust GRP environmental enclosure rated to IP66/7, the analyser is designed for permanent outdoor installation where adverse weather and challenging site conditions prevent the use of standard portable instruments. Integrated thermostatic heating and cooling are used to maintain stable internal conditions for long-term monitoring across a wide ambient temperature range.

Key Features:

  • Gold film sensor technology: Highly specific detection of mercury vapour at trace concentration levels using AMETEK Brookfield’s patented sensor design.
  • Fixed continuous monitoring: Designed for long-term, unattended operation to provide ongoing mercury vapour monitoring.
  • Environmental enclosure: GRP polyester glass-reinforced enclosure rated to IP66/7 for outdoor and harsh environment installation.
  • Integrated temperature control: Thermostatically controlled heating and cooling maintain stable internal operating conditions.
  • Local visual indication: Illuminated display with alarm beacon and health check beacon for immediate on-site status awareness.
  • Detachable handheld analyser: Jerome J405 analyser can be removed from the enclosure for survey work and source identification.
  • Analogue and digital outputs: Passive 4–20 mA output and digital connectivity for integration with site monitoring or control systems.

Have a particular project in mind? We’d welcome a call on +44 (0) 1189 311 188

The system is supplied with a universal backplate and mounting hardware, allowing installation on a wall, pole, fence or custom-built frame. Multiple units can be used to form a perimeter monitoring system for continuous site coverage. An illuminated local display, visual alarm beacon and health check beacon provide clear on-site status indication and support rapid response to elevated mercury levels.

A key operational advantage is the detachable, handheld Jerome J405 Mercury Vapour Analyser within the unit. The analyser can be removed for survey work to help pinpoint the source of mercury contamination so that corrective action can be taken.

The analyser supports both digital and analogue outputs, including a passive 4–20 mA signal for connection to site monitoring or control systems, making it suitable for fixed installation where continuous mercury vapour monitoring is required.

Specifications:

Detection Range0.5 to 999 µg/m³
Resolution0.01 µg/m³ (0.00001 mg/m³)
Accuracy±10% at 1.0 µg/m³; ±5% at 25 µg/m³; ±5% at 100 µg/m³
Response Time16 seconds (sample mode); 2 seconds (survey mode)
Flow Rate750 ± 50 mL/min
Result Unitsµg/m³, mg/m³
Auto Sample Intervals1, 2, 5, 10, 15, 30, 45, 60, 90 or 120 minutes
Optional Data StorageUp to 20,000 test results with date, time and up to 80 locations
Display3.5 inch (9 cm) backlit LCD
Operating Environment-40 to +55 °C (temperature controlled enclosure)
OutputsDigital USB serial output; analogue 4–20 mA passive current loop (±0.3% of output)
EnclosurePolyester glass-reinforced enclosure, IP66 rated
Dimensions600 (H) × 400 (W) × 260 mm (D)
WeightApproximately 12 kg (loaded)
Power Supply110 or 240 VAC site power with internal power supply modules
Summary
Contact

Prefer to speak to a Product Specialist? Call us on:
+44 (0)118 9311 188